Chem 454
April 20
GC- Hardware
Injectors-- gas
- simplest is injecting a gas
- gas syringe, septum, needle
- head space samples (vapor over a liquid)
- trap gases on adsorbant
- heat and desorb into the column
- likely to be 1-2 minutes (too broad)
- cyofocusing
- keep column (column head) cold
- all sample is trapped at head of column; no chromatography
- then, heat to release
- can be -78 C accessory; quick reheat
- can often just be oven at 25oC; temperature program
- specialized units (natural gas) might use valves
Injector-- liquid
- microsyringe (1, 5, 10, 20 microliter sizes)
- quickly inject into hot area
- injector liner (glass sleeve)
- on column injection (if column itself is hot)
- vaporize in <1 sec
- keep injector hot (varies with sample)
-
- packed column can handle a milligram of sample
- 10 microliters (10 mg) at 1% is a mg of sample
- more likely 1-5 microliters, 0.01% or a few mg
- of course, minimum is fixed by detector response
-
- capillary column can handle much less (microgram)
- include a problem with lots of solvent too (can't simply dilute)
- need sample volumes of 0.01-0.1 microliters
- no such syringes
-
- split injectors
- sample vaporizes
- majority of sample escapes
- perhaps 1-5% enters column
- (may also involve timed valves to control sample going to column)
-
Detectors
TCD
- Standard, traditional-- thermal conductivity
- nearly universal, responds to all compounds
- nearly identical response, per mole, for all species
- not terribly sensitive
- rugged, long lived, relatively inexpensive, one gas (He)
-
FID or Flame Ionization Detector
- C- compounds burn in flame (air, H2 added)
- CO2 forms with small fraction as ions
- flame is them electrically conductive
- measure (tiny) electrical currents
- response varies with compound, # carbons
- non-C species often no response
- can detect much tinier amounts than TCD
- also a relatively fast response
- need three gas cylinders
-
- variations (heated ceramic element) are selective for N, P
ECD Electron Capture
- radioactive Beta source
- sample steals electrons, reduces current detected
- very selective
- very halogen sensitive
- often 100-1000 fold less sample than FID
- great with many pesticides, environmental problems
- ignores many other common species
-
Others-- Photoionizeation
GC/MS Definitive Detector
- full mass spectrum for each peak