Chem. 407
electronics experiment


edited: October 18, 2000... document chem407_elect.htm


A few comments on selected portions of the electronics experiment

1. Gain of a simple inverting amplifier


Noise and Filtering

The thermocouple system involves a relatively low voltage signal of a few millivolts. It also involves a relatively long loop of wires that can act as an antenna, picking up small signals of other types. For example, it is easy to pick up 60Hz and 120 Hz signals from the power lines in the room. There are also some other signals generated when electrical wires are flexed or when electrical contacts are pulled or bumped. These undesired extra signals are considered electrical noise. In a typical situation, this noise may be in the neighborhood of a few hundred microvolts.

If the thermocouple signal were large, this noise would be negligible. However, the thermocouple response is in the neighborhood of 500 microvolts per degree C, so the noise represents an uncertainty of 1-2 degrees C.

Fortunately, the interesting thermocouple signal differs from the noise in a substantial manner. The thermocouple signal is steady, changing only slowly over a few seconds or even minutes. The noise changes every few milliseconds. What we need is a frequency sensitive circuit that allows the DC and very low frequency signals to pass but stops or reduces signals of higher frequency. Technically this is called a Low Pass Filter.

We can implement such a filter by adding a capacitor across the feedback resistor of our amplifier. A capacitor acts like a low resistance element for high frequency signals.

A capacitor behaves differently with a DC signal the net result...


the pH meter circuit

This is an example that shows the advantage of taking a step by step approach to design and assembly.

Basically there are four rather distinct features or stages of our pH meter and all must work properly

If any of these components isn't working properly (or connected together properly) the device won't work as intended. There are too many steps to hope we won't have an error.

Instead, let's try to build this circuit in a series of short steps, each with a built in test.

1. Let's test the electrode.

If you don't see a change you either have the meter set wrong, a malfunctioning electrode, bad connections or bad buffers. Find out which before proceeding.

2. Check the voltage follower circuit and the electrode response

If this isn't working, find out what's wrong before doing part 3.

3. Setting up an amplifier with the desired gain

4. Adding a suitable voltage offset so the meter reads pH directly


return to home page